12 Channel ECG machine with interpretation,Real-time ECG waveform,20 cases memory, Built-in USB
ECG-NC12 Interface (ECG software optional), DC/AC power with built-in rechargeable Li-ion battery, 320x240 graphic 6 inch LCD
Features
Features
1. Design patent,software patent
2. Application of CSE database,reliable measurement and analysis results
3. 12-lead simultaneous acquisition
4. 6"LCD,real-time display of 12-lead ECG waveform
5. Detection and alarm on lead-off and low battery
6. Automatic adjustment of baseline for optimal printing
7. Built-in EGC simulator for DEMO purpose
8. Pre-10-second printing to print out any abnormal EGC waveform
9. Unique 6CH+P printing mode,convenient for physical examination and printing
10. report on A4 paper
11. 3kinds of operation mode:AUTO,MAN and ANA
12. 3kinds of filter :HUM,EMG and ADS
13. AC,DC and battery power supply
14. 2 types lead mode:Standard and Cabrera
15. 1,000 EGC files can be saved
16. Communication with PC(Option)
Specification
Safety Standard: | IEC Class 1,type CF |
Acquisition Mode: | Standard 12 lead |
Input Circuit: | Floating,protection against defibrillator and pacemaker |
A/D converter: | 12 bit |
Time Constant: | ≥ 3.2 s |
Frequence Response: | 0.05Hz~165Hz(-3dB) |
Sensitivity : | Auto,2.5,5,10,20,40mm/mV.(±2%) |
Input Impedance: | ≥50MΩ |
Input Circuit Current: | ≤50nA |
Calibration Voltage: | 1mV±2% |
Noise Level: | <15µVp-p |
Anti Baseline Drift : | Automatic |
Filter : | EMG filter 35Hz~45Hz(-3dB) ADS filter:Auto( 0.15~0.50Hz) HUM filter:50Hz/60 Hz(-20dB) |
CMRR: | >100dB |
Patient Current leakage: | <10µA |
Paper Speed: | 6.25/12.5/25/50 mm/s |
Recording Paper: | 210/216 mm roll paper |
Power Supply: | AC:100-240v(50-60hz); DC:19v/3.5A Battery :14.4v /4400mAh,built-in rechargeable lithium battery,over 2 hours working time |
Dimension : | 415mm*195mm*380mm(Carton) |
Net Weight: | 3.5kgs |
Gross Weight: | 5.5kgs |