Main characters
1 It accords with physical engineering principle, 10°inclination design, so the operating feel is more excellent.
2 Air insulation design to avoid cross pollution inside and outside air circulation . 70% Air Recirculation, 30% Air Exhaust,vertical laminar negative pressure.
3 Equipped with spring up/down movable door in the front and back of work bench, flexible and convenient to locate
4 Equipped with special filter on ventilation to keep vented air conforming to national standard.
5 Contact switch adjusts voltage to keep wind speed in working area in ideal state all the time.
6 Operate with LED panel.
7 The material of the work area is 304 stainless steel.
Parameter / model | BSC-1000IIA2 | BSC-1300IIA2 | BSC-1600IIA2 | |
Cleaning level | 100 Level(USA Federal Standard 209E)≥0.5um Granulae≤3.5per liter | |||
Descending wind speed | ≥0.35m/s | |||
Wind speed in | ≥0.55m/s | |||
Noise | ≤58dB(A) | |||
Half ventilation peak value | ≤3µm | |||
Power supply/Max. power consumption | AC220V/50HZ/1KW | |||
Biological safety | Personnel protection: impact type sampler:CFU≤10, | |||
Product protection: all sampling vessels:CFU≤5 | ||||
Cross contamination protection: all doubledishes: CFU≤2 | ||||
working area | 1000*650*580 | 1300*650*580 | 1600*650*580 | |
Overall dimension | 1200*750*1950 | 1500*750*1950 | 1800*750*1950 | |
Specification of air-supply filter | 955*545*50*1 | 1290*554*50*1 | 1590*554*50*1 | |
Specification and number of fluorescent lampor UV lamp | 20W*2/20W*1 | 30W*2/30W*1 | 30W*2/30W*2 | |
Illumination | ≥650LX | ≥650LX | ≥650LX | |
Weight(kg) | 300 | 340 | 380 | |
Suit | Single | Double | Double |