Double Beam Spectrophotometer T80+ with Local & UV Win Software Control

Trade on HiSupplier, Worry Free Guarantee
Trading safety, secure your money, lower the risk, protect both buyers and suppliers
HiSupplier is a Chinese multinational company in U.S, it helps to coordination and order landing.
More ›
  • Product Details
  • Company Profile
  • Product Reviews

 

 

Double Beam Spectrophotometer T80+ with Local & UV Win Software Control

 

T80+ is a high performance double beam spectrophotometer available with a variable spectral bandwidth, which is innovative in terms of instrument application, mechanical and optical design, electronic control and software whilst retaining features that are well established and accepted through the industry.

T80+ is able to carry out photometric measurement, spectrum scans, quantitative determination and DNA/Protein analysis. When interfaced to a PC using the UV-Win software, many more features are available including three dimensional spectrum, kinetic measurements, method and data storage, exportation of data in multiple formats and GLP administration features.


Features & Functions

High performance variable spectral bandwidth
Wavelength accuracy +0.3nm
Supplied with a motorized 8 cell changer and pre-aligned Tungsten and Deuterium lamps
Holographic blazed grating 1200lines/mm
High degree of automation requiring minimal key depressions to start analysis
A number of optional accessories available which increase the flexibility of the analysis
Analysis for photometric measurement, spectrum scans, quantitative determination and DNA/Protein analysis
UV-Win software gives additional functionality including 3D spectrum analysis and compliance with GLP protocol.
Simple mechanical structure and modular electronics make routine maintenance very easy



Applications:

Environmental monitoring & control
Food & Beverage production and quality control
Agricultural monitoring
Bio-tech & life science
Clinical & forensic analysis
Geology & metallurgy
Gas & oil

 

 

 

Specifications:

 

Optical system

Double beam 

Scan speed

Selectable 

Wavelength range

190-1100nm

Wavelength accuracy

+0.3nm

Wavelength reproducibility

≤0.2nm

Spectral bandwidth

0.5, 1.0, 2.0, 5.0nm

Photometric mode

Transmittance, Absorbance, Energy Concentration

Photometric range

-0.3-3.0Abs

Photometric accuracy

0.002A (0-0.5A), 0.004A (0.5-1A), 0.3%T (0-100%T)

Photometric reproducibility

0.001A(0-0.5A), 0.002A(0.5-1A), 0.15%T(0-100%T)

Photometric noise

0.001A(500nm) 30min warm-up

Baseline flatness

0.0015A(200-1000nm)

Baseline stability

0.0008A/h (500nm, 0Abs), 2hr warm-up

Stray light

≤0.12%T(220nm NaI, 340nm NaNO2)

Standard functionality

Photometric, Quantitative, Spectrum and DNA measurements

Cell holder

Automatic 8 cell changer

Detector 

Silicon photo diode

Light source

Tungsten Halogen and Deuterium arc lamps

Display 

Digital 

Send message to this supplier
Ms. April Cao
Tel: 86-10-69910653
Enter between 20 to 3,000 chatacters.
Contact Method
PG Instruments company Limited
No.3 Pingsan Road, Beijing, China (Mainland)
Related Products
Didn't find what you're looking for? Post Buying Lead or contact our customer service specialist for help!