Specifications:
|
Technical data |
WS-150YDC |
WS-200YDC |
WS-280YDC |
WS-400YDC |
|
1 |
Chamber volume(mm) |
150L Φ440x1000 |
200L Φ515x1000 |
280L Φ600x1000 |
400L Φ700x1100 |
|
2 |
Working pressure |
0.22MPa |
|
|||
3 |
Working temperature |
134 |
|
|||
4 |
Adjustment of temperature |
40-134 |
|
|||
5 |
Time for sterilization |
060min |
|
|||
6 |
Timer for drying |
060min |
|
|||
7 |
Heat average |
≤±2 |
|
|||
8 |
Power |
9KW 380V 50HZ |
9KW 380V 50HZ |
12KW 380V 50HZ |
18KW 380V 50HZ |
|
9 |
Dimension (cm) |
140x60x130 |
140x67x165 |
140x77x178 |
143x88x183 |
|
10 |
Transport dimension(cm) |
155x75x185 |
156x82x185 |
168x92x210 |
160x105x210 |
|
11 |
G.W/ N.W |
320/240kg |
350/260kg |
465/365kg |
530/420kg |
|
Characteristic
- Sterilizing course automatic sterilization controlled, easy to operate.
- With drying function, suitable dressing drying.
- With over-temperature, over-pressure auto-protect device.
- The door opening mechanism cannot be operated until the pressure in the chamber is reduced to 0.027Mpa. It can’t be start on if the door doesn’t close well.
- The safety valve will be open automatically when the inner pressure over 0.24Mpa, and the steam be exhausting to the water tank.
- The power will be cut-off automatically, cut-off the water and alarm when the machine lacking water.
- Fully stainless steel structure.
Applications:
The horizontal autoclave, which adopted the way of gravity exchange to releasing the cold air from the chamber more completely to ensure the reliable sterilization.
The control system automatically adjust the steam inlet and outet according to chamber temperature during sterilizing.
This unit is ideal equipment for clinics, scientific research institutes and other organizations to sterilize surgical instruments, fabric glassed, cultrure media etc.