Double Beam UV-Vis Molecular Spectroscopy T80 with 8 Position Sample Holder and Reference

 

 

T80 is a high performance double beam spectrophotometer available with a fixed spectral bandwidth, which is innovative in terms of instrument application, mechanical and optical design, electronic control and software whilst retaining features that are well established and accepted through the industry.

T80 is able to carry out photometric measurement, spectrum scans, quantitative determination and DNA/Protein analysis. When interfaced to a PC using the UV-Win software, many more features are available including three dimensional spectrum, kinetic measurements, method and data storage, exportation of data in multiple formats and GLP administration features.


Features & Functions

 

  • High performance fixed 2nm spectral bandwidth
  • Wavelength accuracy +0.3nm
     
  • Supplied with a motorized 8 cell changer and pre-aligned Tungsten and Deuterium lamps
     
  • Holographic blazed grating 1200lines/mm
     
  • High degree of automation requiring minimal key depressions to start analysis
     
  • A number of optional accessories available which increase the flexibility of the analysis
     
  • Analysis for photometric measurement, spectrum scans, quantitative determination and DNA/Protein analysis
     
  • UV-Win software gives additional functionality including 3D spectrum analysis and compliance with GLP protocol.
     
  • Simple mechanical structure and modular electronics make routine maintenance very easy
     




Applications:

Environmental monitoring & control
Food & Beverage production and quality control
Agricultural monitoring
Bio-tech & life science
Clinical & forensic analysis
Geology & metallurgy
Gas & oil

 

 

Specifications:

 

Optical system

Double beam 

Scan speed

Selectable 

Wavelength range

190-1100nm

Wavelength accuracy

+0.3nm

Wavelength reproducibility

≤0.2nm

Spectral bandwidth

2nm

Photometric mode

Transmittance, Absorbance, Energy Concentration

Photometric range

-0.3-3.0Abs

Photometric accuracy

0.002A (0-0.5A), 0.004A (0.5-1A), 0.3%T (0-100%T)

Photometric reproducibility

00.001A(0-0.5A), 0.002A(0.5-1A), 0.15%T(0-100%T)

Photometric noise

0.001A(500nm) 30min warm-up

Baseline flatness

0.0015A(200-1000nm)

Baseline stability

0.0008A/h (500nm, 0Abs), 2hr warm-up

Stray light

≤0.12%T(220nm NaI, 340nm NaNO2)

Standard functionality

Photometric, Quantitative, Spectrum and DNA measurements

Cell holder

Send message to this supplier
Mr. David Wang (Sale Manager)
Tel: 86-10-69910653
Enter between 20 to 3,000 chatacters.
Contact Method
PG Instruments Limited
No.3 Pingsan Road, Pinggu District, Beijing, China (Mainland) / 101200
Fax: 86-10-69910516
Didn't find what you're looking for? Post Buying Lead or contact our customer service specialist for help!