UV lamp tester
UV lamp tester simulates damaging effects of long term outdoor exposure of materials and coatings by exposing test samples to varying conditions of the most aggressive components of weathering - ultraviolet radiation, moisture and heat. A QUV accelerated weathering chamber uses fluorecent lamps to provide a radiation spectrum centered in the ultraviolet wavelengths. Moisture is provided by forced condensation, and temperature is controlled by heaters.
Model | HD-703 |
The main technical parameters | |
Temperature range | RT+10~70°C |
Humidity range | 75%~95%RH |
Control mode | Balance temperature humidity control (BTHC) |
Distance from sample to lamp: | 50±2mm |
Centre distance between the lamp | 70mm |
Wavelength ultraviolet light | 280~400mm |
Standard sample holder | 75×150×1.5mm |
Shows the value of radiation intensity and totalizer | |
Optional RS-485 communication interface | |
Power | 220v/50Hz /±10% 6KW |
Design standard | ISO-4892, ASTM-G53, D4587, 4799, BS-27282, SAE-J2020 |