- Power entry module with North American style 3AG fuse holder
- 2 or 4 voltage selection
- Compact snap-in design
- Two element circuit provides basic EMI attenuation
- Available with minimal leakage current suitable for medical applications (HJ models)
- Also available without filter (VJ models)
Non-filteredmodels | |
6VJ1 | 6VJ1-2 |
GeneralPurpose Filters | |
6J4 | 6J4-2 |
MedicalFilters | |
6HJ4-4 | 6HJ4-2 |
Specifications
Maximum leakagecurrent, each Line to Ground: | |||
6J4 Models | 6HJ4 or non-filteredmodels | ||
@ 250 VAC 50 Hz | 500 µA | 5 µA | |
Hipot rating (oneminute): | |||
Line to Ground (VAC) | 1550 VAC | ||
Line to Line (VDC) | 1450 VDC | ||
Operating Frequency: | 50 / 60 Hz | ||
Operating voltages: | |||
suffix -1 or -4 models: | 100, 120, 220 or 240VAC | ||
suffix - 2 models: | 115 or 230 VAC | ||
Rated Current: | 2 to 6A | ||
Required Fuse: | .25 x 1.25 (not included) |
Accessories
GA400: NEMA 5-15P to IEC 60320-1 C13 line cord
Self-adhesive, available in multiples of 40, Specify part number
JA302: 2 Voltage Select Card
Comes standard with 6VJ1-2, 6J4-2 and 6HJ4-2
Comes standard with 6VJ1-2, 6J4-2 and 6HJ4-2
JA304: 4 Voltage Select Card
Comes standard with 6VJ1, 6J4 and 6HJ4-4
Comes standard with 6VJ1, 6J4 and 6HJ4-4
JA403: Mounting clips for .105 - .125" panels
JA410-419: Equipment Rating Labels Self-adhesive, available in multiples of 40, Specify part number
Part Number | A VA Max. | B Fuse 100/120(115) | C Fuse 220/240(230) |
JA410 | 25 | 250 mA | 125 mA |
JA411 | 50 | 500 mA | 250 mA |
JA412 | 100 | 1A | 500 mA |
JA413 | 200 | 2A | 1A |
JA414 | 250 | 2.5A | 1.25A |
JA415 | 300 | 3A | 1.5A |
JA416 | 400 | 4A | 2A |
JA417 | 500 | 5A | 2.5A |
JA418 | 600 | 6A | 3A |
JA419 | Assortment, 5 of eachlabel JA410-418 | ||
Label Height: | .210" | ||
Label Width: | 2.330" |
JA500: Voltage Selector Card Extractor Tool
Electrical Schematics
6VJ1 & 6VJ1-2
6J4 & 6J4-2
6VJ1 & 6VJ1-2
6J4 & 6J4-2
6HJ4-4 & 6J4-2
Note 1: Jumper required if only SPST power switch is used
Note 2: Jumpers required if no input filtering is used
Note 3: Use only 120V and 240V positions for 2 volt selection units
Voltage Selection
Note 2: Jumpers required if no input filtering is used
Note 3: Use only 120V and 240V positions for 2 volt selection units
Voltage Selection
Open cover door and slide fuse-pull lever to left. Select operating voltage by orienting voltage selection card with the desired voltage on top left side. Push card firmly into module slot. Slide fuse-pull lever to right into normal position and re-insert fuse into holders.
Use caution in selecting correct fuse value.
Case Styles
Filtered Models
Non-filtered Models
Recommended Panel Cutouts
Standard units mount in panel thickness of .060 - .090 [1.52 -2.29]
JA403 Mounting clips for .105 - .125" panels available separately
Fuse cover door shown in open position
Filtered Models
Non-filtered Models
Typical Dimensions: | ||
Line Inlet (1): | IEC60320-1 C14 | |
Terminals: | .110 [2.79] |
Recommended Panel Cutouts
Standard units mount in panel thickness of .060 - .090 [1.52 -2.29]
JA403 Mounting clips for .105 - .125" panels available separately
Fuse cover door shown in open position
Case Dimensions
Part No. | A | B | C | D |
(max.) | (max.) | (max.) | (max.) | |
Unfiltered 6VJ1, 6VJ1-2 | 2.68 | 1.52 | 1.17 | 1.23 |
68.1 | 38.6 | 29.7 | 31.2 | |
6J4, 6J4-2, 6HJ4-4, 6HJ4-2 | 2.75 | 1.87 | 1.17 | 1.58 |
69.9 | 47.5 | 29.7 | 40.1 | |
Performance Data
Typical Insertion Loss
Measured in closed 50 ohm system
6J4
6HJ4
Minimum Insertion Loss in dB:
Measured in closed 50 ohm system
Common Mode / Asymmetrical (Line to Ground)
Typical Insertion Loss
Measured in closed 50 ohm system
6J4
6HJ4
Minimum Insertion Loss in dB:
Measured in closed 50 ohm system
Common Mode / Asymmetrical (Line to Ground)
Model No. | Frequency - MHz | ||||||
.15 | .5 | 1 | 5 | 10 | 20 | 30 | |
6J4 | 9 | 20 | 25 | 41 | 45 | 45 | 48 |
6HJ4 | 9 | 11 | 15 | 19 | 13 | 12 | 10 |