Features
1.7"TFT Color screen, folding 90 degree;
2.The internal line thermal printer can records FHR ,TOCO ,The life period can exceed over 20years;
5.Auto Fetal movement counting with monitoring curve and data display;
3.1MHZ,9 crystal safer wide beam pulse wave transducer; 4.Low consumption ultrasound transducer, it's safe for fetus monitoring;
5.AC or LI-battery operated;
6.More than 10 hours data storage, then can be reviewed and printed;
7.Build-in communication interface to the central nurse station.
Specification
FHR Measure | Measure Mode | Ultrasonic Doppler |
50~210bpm | ||
Measure Accuracy: | ≤±2bpm | |
FHR Wave Arrange | 50~210bpm | |
FHR Digital Display Arrange | 50~210bpm | |
Strength | <5mW/cm2 | |
Working Frequency | 1.0MHz | |
TOCO Measure | Measure Mode | External Measurement |
Measurement Range | 0-100 units | |
Measure Accuracy | ≤±10% | |
TOCO Curve Range | 0-100 units | |
TOCO Digital Range | 0-100 units | |
Automatic Reset Function | yes | |
Maternal SpO2 Measure | Measurement Range | 70%~99% |
Measure Accuracy | ±3% | |
NIBP Measure | NIBP-SYS | 6.7~32.0kPa(50~240mmHg) |
NIBP-DIA | 3.4~26.6kPa(25~200mmHg) | |
NIBP-MEAN | 2.0~24.0ka(15~180mmHg) | |
NIBP Accuracy | ±1.1kPa(±8mmHg)or ±5% of results, taking the bigger one | |
Measure Mode | Manual/Automatic | |
Pregnant HR Measure | Measurement Range | 40bpm~240bpm |
Measure Accuracy | ±2 bpm | |
Fetal Movement Marker | Icon of fetal movement button in manual Fetal movement detected automatically | |
Event Marker | marked by doctor for clinical event | |
Display Screens | 7" TFT 480*234 | |
Record Mode | thermal dot matrix printer system | |
Record Paper | 112mm folded thermal printer paper | |
Paper Skip Speed | 1\2\3cm/min, accuracy ±5% | |
Fetal Heart Rate Alarming Limit | lower limit, 90, 100, 110, 120 bpm optional. upper limit, 160, 170, 180, 190 bpm optional | |
ALM DELAY | 15 seconds or 30 seconds optional | |
Alarming for printer problems | AL_PAPER: out of paper AL_COVER: Printer panel not closed or not closed completely. | |
Maximum Audio Output Power | <1.5W | |
Power | AC: 100-240V,50/60Hz,20VA | |
DC: 7.4V/2200mAh rechargeable Lithium Battery | ||
Dimension | 295D*240W*73H(mm) | |
Weight | 1.75KG | |
Probe Connecting Line | 2.7~2.95m | |
Network Interface | RS232 ,TIP/IP | |
Working Environment | Temperature: +5℃ ~ +40℃ Humidity: < 85% Atmospheric Pressure: 86kPa ~ 106kPa | |
Storage Temperature: | Temperature: -10℃ ~ +55℃ Humidity: < 93% Atmospheric Pressure: 86kPa ~ 106kPa |
Standard Configuration
1.Three Parameter:FHR,TOCO;AFM/MFM. 2.Six Parameters:FHR,TOCO;AFM/MFM;NIBP,SPO2,MHR.
Optional Function:
1.Twins Function
2.Touch Screen