Overall maximum size: 303.5*138*93mm
Maximum weight: <3.0kg
Power supply: AC100-240V.50/60HZ
Max power consumption: 25VA
Applicable syringe: All kinds of standard syringes: 10ml, 20ml and 50ml
Injection accuracy: ±3%( after being adjusted, up to ±2%)
Bolus rate: 10ml: 100ml/h; 20ml: 200ml/h; 50ml: 500ml/h
Displayed information: injection rate; accumulated injection volume; syringe size; battery capacity; battery charging indication; AC power supply connection indication; BED NO. alarm information
Injection increment: 0.1ml/h
Preset injection range setting: 0.1~999.9ml
Accumulated injection volume display:0.1~999.9ml
Battery: Rechargeable lithium polymer battery 7.4V, 1600mA Working for 4 hours (25ml/h) after fully being charged
Electromagnetic compatibility:Meet the requirements of IEC60601-1-2(2001 version)
Electrical safety: Meet the requirements of IEC60601-1:1995, GB9706.1-2007
Injection rate range setting:10ml: 0.1~100ml/h; 20ml: 0.1~200ml/h; 50ml: 0.1~500ml/h
Alarm content: over; injection near completion; lower battery; disengaged; pump stuck; CPU communications error