(1)Features:
1.With 3 1/2 LED,display is clear and precision is high.
2.Electrode constant compensation.
3.Manual temperature compensation.
4.Output signal (0-10)mV
1.With 3 1/2 LED,display is clear and precision is high.
2.Electrode constant compensation.
3.Manual temperature compensation.
4.Output signal (0-10)mV
(2)Technical parameters:
1. Measuring range:
The measuring range is (0~2*10^5) uS/cm, the instrument has five ranges and can be changed through band switch by manual (by electrode with constant).
Range Types | Measurement Range | Resolution |
2μS/cm | (0.001~2)μS/cm | 0.001μS/cm |
20μS/cm | (0.01~20)μS/cm | 0.01μS/cm |
200μS/cm | (0.1~200)μS/cm | 0.1μS/cm |
2mS/cm | (0.001~2)μS/cm | 0.001mS/cm |
20mS/cm | (0.01~20)μS/cm | 0.01mS/cm |
NOTE:To measure high conductivity ,a larger constant electrode is commonly used. If the conductivity is over 10000 uS/cm,the electrode with constant of 10cm-1 is used and the measuring range is extended to 2*105 uS/cm.