 Automatic flash point apparatus Conforms to standards:

1.ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester

2.GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method .

 

Summary

Automatic PMCC Flash Point Tester determines flash points of a wide range of products by a closed cup method with two option speed stirring of the sample. Extensively used in shipping and safety regulations for detection of contamination by volatile and flammable materials in fuel oils and lubricating oils, and for characterization of hazardous waste samples.

 

I. Main Technical Features

Ø 1. Simulation tracking displays function curves of heating up and test time.

Ø 2. Automatic Flash Point Tester Equips with the English prompt function of mis-operation.

Ø 3. Automatic Flash Point Tester Equips with prompt function of test date,test time and other parameters.

Ø 4. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.

Ø 5. Differential coefficient detection.Automatically correct the system deviation.

Ø 6. Automatically open the lid,detect the flash point and print test data.

Ø 7. The test arm rises up and lowers down automatically.

Ø 8. Igniting with electric igniting gun,forced air cooling.

II. Main Technical Specifications

1. Power supply:AC(220±10%)V,  50Hz.

2. Temperature measurement:  Range: Room temperature~300℃.

                                              Repeatability: ≤ 2℃

                                              Reproducibility: ≤ 4℃

                                              Accuracy: 0.1℃

3. Heating rate of Automatic Flash Point Tester

                  Procedure A: (5~6)℃/min

                        Procedure B: (1~1.6)℃/min

                        Automatic control and manually adjustable.

4. Stirring rate:(1) Procedure A: (90~120)RPM

                        (2) Procedure B: (250±10)RPM

                        Automatic control and manually adjustable.

5. Igniting mode: (1) Igniting source: gas(or other civilian fuels,the same below)

                         (2) Electric ignition. Gas flame diameter:3.2mm~4.8mm

6. Ambient temperature:≤ 40℃

7. Relative humidity:≤ 80%

8. Maximum power consumption:350W


Send message to this supplier
Mr. lu
Tel: 86-23-62940053
Enter between 20 to 3,000 chatacters.
Contact Method
chongqing gold mechanical and electrical equipment co.,ltd
9F/10 Jintai building nanping street nanan district, Chongqing, China (Mainland)
Didn't find what you're looking for? Post Buying Lead or contact our customer service specialist for help!
Buyer Help Center