Adown Sensor Probe Carbon Monoxide Tester With Relay Output And Modbus

Trade on HiSupplier, Worry Free Guarantee
Trading safety, secure your money, lower the risk, protect both buyers and suppliers
HiSupplier is a Chinese multinational company in U.S, it helps to coordination and order landing.
More ›
  • Product Details
  • Company Profile
  • Product Reviews

Specification
 

Sensors

Gas Sensor

Electrochemical carbon monoxide sensor

Sensor lifetime

Typically more than 3 years, replaceable

Warm up time

60 minutes (for the first time use)   

Response Time

Within 60 seconds

Signal Update

1s

CO Measuring Range

0~100ppm(default)/0~200ppm/0~500ppm  selectable

Accuracy

<1ppm+5%reading

Stability

±5% (over 900 days)

Temperature & Humidity Sensor (optional)

Temperature

Relative Humidity

Sensing element:

Band-gap-senor

Capacitive humidity sensor

Measuring range

-10~60

0 -100%RH

Accuracy

±0.5 (20~40)

±4.0%RH (25, 15%-85%RH)

Display resolution

0.1

0.1%RH

Stability

±0.1 per year

±1%RH per year

 
 

Outputs

LCD Display (optional)

Display real time CO measurement
or CO+ temperature& humidity measurements

Analog Output

1X010VDC or 4~20mA linear output for CO measurement

Analog Output Resolution

16Bit

Relay dry contact Output

Up to two dry-contact outputs
Max, switching current 3A (230VAC/30VDC), resistance Load

RS485 communication interface

Optional Modbus RTU protocol with 19200bps(default),
Or BACnet MS/TP protocol with 38400bps(default)

 
 

Electrical and General Items

Power Supply

24VAC/VDC                

Power Consumption

2.8W

Wiring Standard

Wire section area<1.5mm2

Working Condition

-10~60(14~140); 5~99%RH, non condensing

Storage Conditions

-10~60(14~140)/ 5~99%RH, non condensing

Net Weight

260g    

Manufacturing Process

ISO 9001 Certified

Housing and IP class

PC/ABS fireproof plastic material, protection class: IP30

Compliance

CE-EMC Approval

 
 
Features

  • Design for real-time detection air carbon monoxide.
  • High accuracy humidity and temperature detection optional
  • LCD display carbon monoxide and optional temperature & RH measurement.
  • Smart buttons for easy operation
  • Excellent electrochemical CO sensor with more than 3 years lift time in typical use
  • CO sensor is replaced
  • Provide 1X analog linear output (0~10VDC/4~20mA selectable) for the measurement
  • Providing up to two dry contact outputs which are controller the setpoint
  • RS485 Modbus /BACnet interface optional
  • 24VAC/VDC power supply
  • CE-Approval

 Typical Application
 

  1. In undergroung parking lots and garages to detect CO and control ventilators
  2. In offices and public places to detect and control CO concentration
  3. In BAS to detect CO concentration 
  4. For all ventilation control systems

Model guide
 
TKG- CO- 1 X Y Z C - R–TH/T
 

1:  wall mounted with the external sensor in an adown probe
 
X: analog output
0: no analog output
1: 1x0~10VDC linear output(default)
2: 1X4~20mA linear output (default)
3: 1X0~10VDC output and can be switch to 4~20mA (just for only one analog output)
 
Y: relay dry contact output
0: no dry contact output
1: 1xdry contact output
2: 2xdry contact outputs
 
Z: Communication interface
0: no communication interface
1: Modbus RS485
9: BACnet communication
 
C: 24VDC/VAC power supply 
 
R: co range
001: 0~100ppm (default)
002: 0~200ppm
005: 0~500ppm
 
-TH(option)  temperature & RH detection and display
No TH indicates without Temp & RH detection and display
T- Temp detection and display
NL option- No LCD display, only for CO detect without relay output
 
For example: TKG-CO-1101-001-TH indicates the CO detector with 1X0~10VDC linear output and Modbus interface. CO range:0~100ppm. It has temperature and humidity detection and display.
 
Demmensions & Mounting

Wiring diagram
 

Pictures

Send message to this supplier
Mr. Tian Qing
Tel: 86-10-59738930
Enter between 20 to 3,000 chatacters.
Contact Method
Tongdy Control Technology Co., Ltd
Building #8, Courtyard #9, Dijin Rd., Beijing, China (Mainland)
Didn't find what you're looking for? Post Buying Lead or contact our customer service specialist for help!