Motorized Hinge Door Medical Autoclave Sterilizer For Pharmaceutical Industry
Features:
- Ideal steam sterilizers for dealing with rubber plugs and aluminum caps in pharmaceutical industry
- High capacity and strong special post vacuum drying function
- Touch screen technology for high user interaction
- Possibility to have the technical area for easy maintain on the left or on the right or both sides
- High effective liquid-ring vacuum pump for saving energy
- Excellent ergonomics together with well-adapted accessories
- Integrated automatic PLC control system with one key operation function
- High quality chamber with fine polished of quality stainless steel
Specification:
Model |
MDB-0.36 |
|
Style |
Single Door (S) |
Double Door (D) |
Overall Dimension |
1500*1800*1340mm |
1500*1800*1340mm |
Chamber |
600*600*1000mm ,Vol: 360L, Material: STS 316L |
|
Net weight |
800kg |
|
Loading System |
Loading Car,Loading Shelf |
|
Operating Temperature |
105°C to 134°C
|
|
Air Removal |
Vacuum / Gravity |
|
Controls |
Microprocessor |
|
Display |
5.7inch color touch screen |
|
Printer |
Built-in printer |
|
Power |
Our ordinary electrical power is AC380, 50Hz, 3phase, N, E
|
|
Steam feeding |
External steam supply, 0.4---0.7MPa, saturated steam offer by user Built-in electrically heated steam generator (Chamber volume <= 600L) Built-in industrial steam heated steam generator (Chamber volume <= 600L) Combining steam generators: individual electrically heated steam generators or natural gas heated steam generators
|
|
Water |
Softened water (Hardness <50ppm/pH Neutral) is at least requirement for the autoclaves’ durability and safety of the goods to be sterilized. Water pressure: 0.15----0.3MPa
|
|
Compressed air |
Oil free compressed air is important to the autoclave’s running. Oil free silence air-compressor is optional for user’s selection. Compressed air pressure: 0.5---0.8MPa
|
Sterilization Cycles:
Cycle No. |
Cycle Name |
Further Details |
1 |
Rubber plugs |
3 times Vacuum, expose temperature 121, expose time 25 min, drying time 25 min. |
2 |
Aluminum caps |
3 times Vacuum, expose temperature 132, expose time 15 min, drying time 15 min. |
3 |
Clothing |
3 times Vacuum, expose temperature 121, expose time 15 min, drying time 25 min. |
4 |
Liquid |
15 min exchange, expose temperature 121, expose time 15 min, cooling time 25 min. |
5 |
B-D Test |
3 times Vacuum, expose temperature 134, expose time 4 min, drying time 3 min. |
6 |
Vacuum Leakage Test |
Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. |
7 |
User Programmable |
All preset parameters can be resettable; Multiple customized programs can be set and stored by user. |
Specialized In Steam Sterilization:
From the beginning as a sterilizer service company, a deep understanding of the sterilization process has been essential to our growth and success as a sterilizer manufacturer.
We utilize this understanding of sterilization to design, engineer and manufacture the best sterilizer available in the market.
Every sterilizer we build is designed, manufactured, assembled and tested at our company headquarters in China . Sterilizer assembly and testing is completed in our state-of-the-art stainless and carbon steel only facility.